Battery External Short Circuit Testing Equipment With Current Temperaure Voltage
Product Details:
Place of Origin: | Guangdong, China |
Brand Name: | GAOXIN |
Certification: | ISO 2008 |
Model Number: | GX-6055 |
Payment & Shipping Terms:
Minimum Order Quantity: | 1set |
---|---|
Price: | negoatition |
Packaging Details: | Non-fumigation wooden box |
Delivery Time: | 15 work days |
Payment Terms: | T/T |
Supply Ability: | 20 sets per month |
Detail Information |
|||
Material: | 304# Stainless Steel | Impact Voltage: | AC 1kv/1.2-50us(peak)1min |
---|---|---|---|
Maximum Short Circuit Current: | 1000A | Action Time: | Response Time/release Time≯30ms |
Mechanical Life: | 30 Thousand Times | Temperature Range: | +10~100℃ |
Test Hole: | Diameter Of 50 Mm | ||
High Light: | battery testing instruments,battery testing machine |
Product Description
Working Ambient Temperature 10 ℃ ~ 100 ℃ Battery External Short Circuit Testing Equipment
1. Sepecifications
Voltage,temperature, current collection | External connected multi - channel data acquisition instrument or PLC data collection module, Channels are extended in Nx4 mode |
Explosion-proof temperature Range | RT + 10~200 ℃ |
Heat preservation material | Efficient compression glass wool or Rigid polyurethane foam cotton |
Temperature control accurcey | 0.1℃ |
Temperature uniformity | ± 2 ℃ |
Temperature luctuation | ± 0.5 ℃ |
Explosion inner box size | 500*500*500mm, |
Short circuit Mode |
RT Short circuit high temperature short circuit |
Operating ambient temperature | temperature 10℃~100℃, humidity 10~90% no condensation |
Power | AC 220V± 10%, 50/60Hz, 380V ± 1% |
Gas consumption | 0.5~0.8MPa |
Working voltage | AC 220V 50Hz-60Hz |
Impact voltage | AC 1kv/1.2-50μs(max)1min |
Maximum testing voltage | 100V~1000V |
Dc response time | ≤5μs |
Maximum short circuit current | 500A/1000A |
The device interior resistance | 80±20mΩ |
Action time | Response time / release time ≯30ms |
Performance characteristic |
Cold and voltage ≯66%Us Cold release voltage ≯30%Us,≮5%Us |
Operating Mode | The remote control with wire or Remote control within 7 meters without barrier |
Explosion-proof case material |
Inner case: SUS# 304 stainless steel sheet Outer case: Cold-roll steel sheet with paint |
Cotrol Mode | Touch-tone or PLC touch screen full-automatic control |
2. Description
Remote controlling large current and short circuit device is designed by combining various battery short circuit test standard requirements. It conforms to such standards and codes as GB/IEC/EN/UN38.3/UL; According to the standard requirements, the short circuit device must conforms to the internal resistance range far less than 100mΩ so as to achieve the maximum short circuit current required by the test.
3. Corresponding Stardards
GB 18287-2013, QB/T 2502-2000, IEC 62133-2012, UN38.3, UL 2054, UL 1642, QC/T 744-2006, QC/T 743-2006, IEC 62281-2004, GB 8897.4-2002, IEC 60086-4:2000, SNT 1413-2004, SNT 1414.3-2004